12 INCH PATIENT MONITOR PM-LP12A NIBP , SPO2 , TEMP, RESP, ECG, ETC02 TECHNICAL SPECIFICATIONS Multi-parameter Patient Monitor

  • CE, ISO13485
  • SFDA: Class ���b
  • Anti-electroshock degree: Class���equipment (internal power supply)
  • TEMP/SpO 2 /NIBP: BF
  • ECG/Resp: CF Quality Standards and Classification
  • Adult/Pediatric/Neonatal/Medicine /Surgery/Operating Room/ICU/CCU Application Range
  • 12��� realcolorTFTscreen Resolution:800��600
  • One alarm indicator(yellow/red)
  • One working indicator(green)
  • One battery charge stateindicator(green)
  • Three modes in accordance with the alarm state Display Operating Environment :
  • Temperature: 0 ~ 40 ���
  • Humidity: ���85%
  • Altitude: -500~4600m Transport and Storage Environment :
  • Temperature: -20~60 ���
  • Humidity: ���93%
  • Altitude: -500~13100m Environment
  • AC: 100 ~ 240V, 50Hz/60Hz
  • DC: Built-in rechargeable battery
  • Battery: 5200mAh 11.1V lithium battery 4h operating afterf ull charge(one piece) 5min operating after low battery alarm Power Requirements
  • Equipment: 300mm��286mm��155 mm(about); 4.5kg(about)
  • Package : 350mm ��380 mm �� 300mm(about); 6.5kg(about) Dimension and Weight
  • Trend diagram/table: 7x24h
  • NIBP review: 400 events
  • Wave review: 100min
  • Alarm review: 100 alarm events
  • Support drug concentration titration analysis Date Storage
  • Range: 0 ~ 100%
  • Resolution: 1%
  • Precision: 70% ~ 100%: ��2 DIGIT 0% ~ 69%: ��no definition given SpO2
  • Range: 20 ~ 300bpm
  • Resolution: 1bpm
  • Precision: ��3bpm Pulse Rate
  • Channel 2
  • Measure & alarm range: 0 ~ 50 ���
  • Resolution: 0.1 ���
  • Precision (no sensor): ��0.1 ��� TEMP
  • Etco2 (Optional)
  • IBP modular (Optional)
  • Touch screen (Optional)
  • Trolley bracket (Optional)
  • Hanging bracket (Optional)
  • Monitor recorder (Optional)
  • Wifi (Optional)
  • C.O.modular
  • BIS modular Optional
  • 5 Leads: RA, LA, LL, RL, V
  • Lead mode: ���,���, ���, aVR, aVL, aVF, V Increase: ��250�����500�����1000��� ��2000
  • Sensitivity: > 200 uV (Peak-to-peak value)
  • Input impedance: > 5 ( megohm )
  • Bandwidth: Surgery 1 ~ 20 Hz
  • Monitor 0.5 ~ 40 Hz
  • Diagnostic 0.05 ~ 130 Hz
  • CMRR 100dB
  • Polarization Voltage: �� 300mV
  • Baseline Recovering Time: After defibrillation< 3 seconds
  • Signal Range: 8 mV ���Peak-to-peak value���
  • Calibrating Signal: 1mV, precision ��5%
  • Heart rate range���15-380bmp
  • Heart rate accuracy:+/-1% ECG
  • Method: RA-LL impedance
  • Resp Impedance range: 0.3 ~ 3
  • Base Impedance range: 200 ~ 4000
  • Bandwidth: 0.1 ~ 2.5Hz
  • Resp rate: Adult 0 ~ 120BrPM
  • Neonatal/Pediatric 0 ~ 150BrPM
  • Resolution: 1BrPM
  • Precision: ��2 BrPM
  • Asphyxia Alarm: 10���40 seconds RESP
  • Method: Pulse wave oscillometry
  • Work mode: Manual/ Auto/ STAT
  • Measure interval of auto mode: 1,2,3,4,5,10,15,30,60,90,120,180, 240,480 minute(s)
  • Measuring Time of STAT Mode: 5 minutes
  • PR range: 30 ~ 250bpm
  • Measure & alarm range: Adult SYS 40 ~ 270mmHg DIA 10 ~ 215mmHg MEAN 20 ~ 235mmHg Pediatric SYS 40 ~ 200mmHg DIA 10 ~ 150mmHg MEAN 20 ~ 165mmHg Neonatal SYS 40 ~ 135mmHg DIA 10 ~ 100mmHg MEAN 20 ~ 110mmHg
  • Static pressure range: 0 ~ 300mmHg Precision: �� 3mmHg
  • Pressure precision: Max. average error: ��5mmHg Max. standard deviation: ��8mmHg
  • Overvoltage protection: Adult 300mmHg Pediatric 240mmHg Neonatal 150mmHg NIBP
  • Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
  • Measuring and alarm range
  • ART 0 ~ 300 mmHg
  • PA -6 ~ 120 mmHg
  • CVP/RAP/LAP/ICP -10 ~ 40 mmHg
  • P1/P2 -10 ~ 300 mmHg
  • Press Sensor
  • Sensitivity 5 uV/V/mmHg
  • Impedance 300-3000
  • Resolution: 1 mmHg
  • Accuracy: ��2% or ��1mmHg, which great
  • Actualization interval: about 1 Sec. IBP
  • Side/Main stream
  • Warm-up time:when the ambient temperature is 25 ���, the carbon dioxide curve(capnogram) can be displayed within 20/15 seconds, and all the specifications can be fulfilled within 2 minutes.
  • Measurement range:0-150mmHg, 0-19.7%,0-20kPa (at 760mmHg), atmospheric pressure provided by the host.
  • Resolution 0.1mmHg ���0-69mmHg 0.25mmHg��� 70-150mmHg
  • Precision 0-40mmHg��� ��2mmHg 41-70mmHg��� ��5%���reading��� 71-100mmHg��� ��8%���reading��� 101-150mmHg�����10%���reading���
  • Respiratory rate range 0- 150 BPM
  • Respiratory rate accuracy: ��1 BPM CO2
  • NIBP cuff & tube
  • ECG cable & electrodes
  • SpO2 sensor
  • TEMP probe
  • Lithium-ion battery
  • Power cable
  • Operator���s manual